Week 7 HW: Measurement and analytical tools (Waters)
1. A tool used earlier in the class: the fluorescence plate reader
a. Sample introduction. Manual or robotic pipetting of a few tens of microliters per well into a 96 or 384 well microplate. The plate is loaded on a moving stage and each well is read in turn. No separation step: whatever is in the well is read as a single bulk sample.
b. Physicochemical property. Absorption and re-emission of light by a fluorophore. For GFP and its variants the chromophore is formed autocatalytically inside the folded beta barrel, so fluorescence reports on correctly folded protein, not just on protein mass. In absorbance mode (OD600) the property is light scattering by cells.
c. What it measures. Relative fluorescence intensity per well, proportional to the concentration of mature fluorophore. With a purified standard curve this converts to protein concentration. Kinetic reads give rates, which is how a cell free reaction yield or an enzyme assay is followed over time.
d. Detection. A lamp or LED plus an excitation filter or monochromator illuminates the well, emitted light passes an emission filter and lands on a photomultiplier tube or photodiode, which is digitized into arbitrary fluorescence units.
e. Limitations.
- No molecular identity. The reader reports that something in the well emits at the chosen wavelength, not what it is. A truncated or misincorporated variant that still folds its barrel gives the same signal.
- Requires a reporter. A protein that is not intrinsically fluorescent, such as PETase or Reteplase, needs a separate functional assay or a fusion tag that may itself perturb the protein.
- Bulk measurement with no separation, so background from lysate autofluorescence, media components or scattering is included in the signal.
- Dynamic range problems: inner filter effects and detector saturation at high concentration, poor signal to noise at low concentration.
- Says nothing about mass, sequence, post-translational modifications or folding state beyond the single yes/no of chromophore maturation.
2. LC-MS
a. Sample introduction. Into the LC: an autosampler draws a few microliters from a vial and injects it into the mobile phase stream, which a high pressure pump drives through a column packed with coated silica particles. Into the MS: the column effluent flows straight into the source. A metal capillary held at high voltage sprays the liquid as charged droplets (electrospray ionization). Heat and nebulizing gas evaporate solvent until Coulomb fission leaves bare, multiply charged ions in the gas phase. Other introduction routes exist: direct infusion from a syringe pump, which is what the CDMS used, and DESI for imaging a tissue section.
b. Physicochemical properties. Two stages, two properties.
- LC: differential partitioning between a mobile phase and a stationary phase. Depending on column chemistry this is hydrophobicity (reversed phase), charge (ion exchange), isoelectric point, or hydrodynamic size (size exclusion). Analytes elute at different retention times.
- MS: mass to charge ratio in an electric field. In a time of flight analyzer a fixed potential gives every ion the same kinetic energy, so lighter ions fly faster down a field free tube and arrive first. If an ion mobility cell is present, a third property is added: collision cross section, which is the shape of the ion as it is slowed by collisions with a buffer gas.
c. What it measures. Retention time plus accurate mass. Together these give identity and purity: intact molecular weight of a protein, the mass shift of a post-translational modification, the presence of fragments, aggregates or contaminants, and quantitation against a standard. Fragmenting ions in a collision cell and reading the fragment masses gives sequence information, which localizes a modification or a point mutation to a specific residue.
d. Detection. Optionally a UV detector after the column. In the MS, an electron multiplier: the ion strikes an electron rich surface held at opposite potential, releasing a cascade of secondary electrons that amplifies down an array of channels into a measurable current, which is digitized as intensity versus flight time. Flight time converts to m/z, and m/z converts to mass once the charge state is assigned from the spacing of the charge envelope.
e. Limitations.
- The analyte must ionize. Poorly ionizing species are invisible, and in a complex mixture abundant species suppress the ionization of rare ones.
- The instrument measures m/z, not mass. Multiply charged protein spectra must be deconvoluted, and overlapping envelopes from several species complicate that.
- Mass ceiling: a TOF struggles above roughly 900 kDa, which is why viral capsids and lipid nanoparticles need a charge detection instrument instead.
- Buffer incompatibility. Salts, detergents and glycerol suppress signal, so samples need desalting or buffer exchange; native work requires volatile buffers such as ammonium acetate.
- Destructive, and each sample takes minutes rather than the seconds of a plate read, so it is not a primary high throughput screen.
- Lower structural resolution than crystallography or NMR: ion mobility gives a collision cross section, not atomic coordinates.
- A targeted instrument such as the TQ Absolute only sees what you tell it to look for, so unexpected impurities are missed.
3. Two instruments from the lab tour
BioAccord (LC-TOF)
a. What makes it unique. A compact, self-diagnosing LC-MS built for users who are not mass spectrometrists, with routine characterization as its target: intact protein mass, peptide mapping, glycan profiling. Low to mid sensitivity, but robust enough for process analytics in manufacturing.
b. Output. A chromatogram (intensity versus retention time), and for each peak a mass spectrum of m/z versus intensity. A protein gives a charge state envelope of many peaks that deconvolutes to a single molecular weight. A single ion can be plotted over time as an extracted ion chromatogram.
c. What I would measure. From HW3: whether the cell free reaction actually made Reteplase, and whether its nine disulfide bonds formed. Intact mass confirms the full length chain (expected around 39 kDa) and rules out truncation. Comparing the intact mass of the non-reduced sample against the same sample after DTT reduction gives the key number: every disulfide bond formed removes two hydrogens, so a fully oxidized nine bond protein is 18 Da lighter than the fully reduced one. That is a direct readout of whether the oxidizing/DSB regime worked, which an activity assay only reports indirectly. A tryptic peptide map on the same instrument would then tell me which cysteines are paired.
d. Sample format. Two preparations. For intact mass, denatured: the reaction diluted into acidified water/acetonitrile, desalted away from the lysate buffer, split into a non-reduced aliquot and a DTT reduced one. For the peptide map, denatured, reduced, alkylated with iodoacetamide and digested with trypsin. Note that iodoacetamide is already part of the Reteplase lysate preparation in HW3, so free cysteine status has to be interpreted with that in mind.
Cyclic IMS (QTOF with integrated ion mobility)
a. What makes it unique. Ion mobility is built into the flight path, so ions are separated by shape as well as by m/z. Combined with the gentleness of electrospray, which can carry a protein into the gas phase still folded, it probes secondary and tertiary structure and separates species of identical mass but different conformation. It also accepts varied sample introduction, including DESI for 2D imaging.
b. Output. Three dimensional data: a mobilogram in mobility space, a chromatogram, and a mass spectrum, which can be sliced in any combination. Collision cross section values per charge state. In imaging mode, an image of ion intensity as a function of position on a tissue section.
c. What I would measure. From my final project: whether the beta-1,3-glucanase is actually folded after export. This is open question 2 of my project. The native enzyme is exported by the Tat pathway, which exports already folded protein, and my construct forces it through Sec, which exports unfolded chains that must fold outside the cell; it also carries four cysteines in a Gram positive host with poor disulfide machinery. Native ion mobility on the released enzyme gives a collision cross section distribution. A tight, compact distribution matching a correctly folded reference means the Sec route works. A broad distribution shifted to larger cross sections means a partly unfolded or heterogeneous population, which would explain a low laminarin activity result and would push me toward a Tat signal peptide instead.
d. Sample format. Native. The glucanase released into culture supernatant, concentrated and buffer exchanged into ammonium acetate, with no detergent, no glycerol and no non-volatile salts. Because ion mobility compares conformers rather than giving absolute structure, I would run a denatured control (acid or organic solvent) to anchor the unfolded end of the scale, and ideally the same enzyme produced recombinantly and verified active as the folded reference.
4. Measuring something for the final project
The project. Lactiplantibacillus plantarum engineered to display a beta-1,3-glucanase on its surface through a LysM anchor, released only when Candida albicans secretes its aspartyl protease Sap2, which cuts an engineered linker. The single most important unresolved question is open question 1: does Sap2 actually cut that linker, and does it cut at the designed position?
Instrument. BioAccord for the primary readout, with the Xevo QTOF if sequence level confirmation is needed.
Sample preparation. Grow the engineered strain and a non-cleavable linker control strain. Split each into three conditions: buffer only, purified recombinant Sap2, and C. albicans conditioned medium. Incubate, pellet the cells, and take the supernatant, which should contain the released enzyme only in the cleaved conditions. Concentrate and desalt the supernatant, then run it two ways: intact (denatured, reversed phase LC-MS) for the mass of the released species, and reduced, alkylated and trypsin digested for peptide mapping on the QTOF.
Data generated. A chromatogram with a peak for the released protein, a charge envelope deconvoluting to one molecular weight, and from the digest a set of peptide masses with fragment spectra covering the linker region.
What it tells me. The intact mass identifies the cleavage site directly. The designed cut is at KLVEL / LYLVK, which predicts a released species beginning at Leu and carrying the glucanase plus the C-terminal half of the linker, an estimated 45 to 47 kDa from residue count. A mass a few hundred daltons off that value means Sap2 cut somewhere else in the linker; a mass matching the whole mature fusion including LysM means the anchor fell off the cell wall rather than the linker being cut, which is a different failure and a different fix. The peptide map resolves the ambiguity completely by giving the exact N-terminal peptide of the released protein. The control strain tells me how much release is background leakage from the non-covalent LysM anchor rather than Sap2 dependent, which matters because that leakage is also my fallback mechanism for C. glabrata, which secretes no characterized Sap.
How it advances the project. The whole design rests on a proteolytic switch that has never been tested inside an anchored fusion. A Western blot would only tell me that something of roughly the right size appeared in the supernatant. The mass spectrometer tells me exactly which bond was cut and how much of the release is actually triggered, which is what decides whether the linker stays as designed, gets a different P1 residue, or needs to be replaced.