Week 7 HW: Mass Spectrometry

My proposed burn scaffold uses one protein with two parts: one collagen-like and one elastin-like. How would I check that the two parts are joined?

1. A familiar tool: the fluorescence plate reader

Each liquid sample goes into its own well. The reader shines blue light on sfGFP, our green fluorescent protein. A filter lets the green light it emits reach a detector.

I would subtract the background signal from a well without sfGFP DNA, then compare my recipe with less coenzyme A to the original. A brighter well suggests more fluorescent sfGFP. Known concentrations would let me estimate the amount. The reader misses protein that does not glow.

2. How LC-MS works

A liquid sample is injected onto a column. In reversed-phase LC, molecules that stick more strongly to the column’s water-repelling surface come out later. At the outlet, electrospray makes charged droplets. As they dry, ions enter the MS. A time-of-flight instrument measures when the ions hit a detector to determine their mass-to-charge ratio (m/z).

A chromatogram plots signal against time. A mass spectrum plots signal against m/z. These help identify molecules in a mixture, but two different molecules can have the same mass.

3. Two instruments from the Waters tour

InstrumentSample I would prepareWhat I would learn
BioAccord: whole-protein massPurified sfGFP with salts removed, unfolded but still in one piece.Software combines the charge peaks in the mass spectrum to estimate the protein’s mass. Even a dim sample can be checked for a protein of the expected mass.
Xevo QTOF: peptide mappingScaffold protein cut into short pieces (peptides), then desalted.The instrument selects a peptide ion and breaks it into fragments. Their mass spectrum helps me identify which part of my design the peptide matches.

4. My project: checking the scaffold before making a sheet

I would run a buffer blank, then a purified, desalted scaffold sample on BioAccord. I would compare its measured mass with the mass predicted from my design. A lower mass could mean part of the protein is missing.

On a separate digest, I would use Xevo QTOF to look for a peptide crossing the join, if the digest produces one. Pieces from each half could come from two separate proteins; a piece across the join is evidence they were connected. My next test would be whether a sheet made from the protein holds together when wet. A mass spectrum cannot tell me that.

AI use: I used Codex to discuss the instruments and edit this write-up.