Week 7 HW: Measurement & Assays
Goal: For this homework you will reflect on the different analytical tools you have used in class so far and integrate the learnings from the analytical module.
Analytical technologies are designed to measure different aspects of molecules and samples. All analyzers include the following components: 1) a mechanism to introduce the sample, 2) a mechanism to probe the sample, leveraging physiochemical properties of the target analytes, 3) a mechanism to detect the probed response (i.e. a detector/digitizer), and 4) a mechanism to analyze the data. In this module, liquid chromatography and mass spectrometry were introduced as advanced tools for quantifying and characterizing the structure of large and small molecules. Mass spectrometers serve to separate and detect ions by their mass-to-charge (m/z) ratios, allowing identification by molecular weight. Virtually every analytical technique performs better with improved purity of sample; liquid chromatography can facilitate the specific detection of molecules by separating them in the time domain (prior to UV or MS detection) using properties such as the differential hydrophobicity, isoelectric point, or size of the molecules.
1. For one of the analytical tools you have worked with thus far (e.g. plate readers, gel electrophoresis), describe:
[Gel Electrophoresis]
a. How is the sample introduced?
Sample is introduced by pipetting in a certain amount of liquid into the electrophoresis well.
b. What physiochemical property is being leveraged to study the sample with this tool?
The electric charge of DNA molecules.
c. What does this allow you to measure?
Assuming DNA Molecules are uniformly negatively charged, and the gel is uniformly dense, the gel electrophoresis separates each molecules based on its size, as bigger molecules are held back while smaller fragments are carried further. Some variations in DNA such as baterial plasmid/chromosomes supercoiling could affect this results but generally size are correlated with travel distance.
d. How is the output signal detected?
Signal is detected by using DNA binding dye (Loading dye) and UV reflecting dye on the gel (traditionally EtBr or more recently the safer Fluorosafe for example). Results are then visualized on a UV visualizer.
e. What limitations does the tool have?
- Single molecule type only, so RNA needed to be converted into its cDNA, and protein would need a different dye and gel altogether
- Some toxic reagents which have slowly been replaced by safer ones
- Dependent on DNA extraction works and initial DNA in the cell to have good result
2. For a liquid chromatography mass spectrometry (LC-MS) instrument:
a. How is the sample introduced into the LC? Into the MS?
Into the LC : using a high pressure pump to move the sample through the chromatography column Into the MS : ionizer and sprayer
b. What physiochemical properties are being leveraged to study the sample with this tool?
LC : molecular size and property (depending on which column is used) MS : molecular mass
c. What does this allow you to measure?
Allows for the measurement of the components of a liquid mixture, such as cellular lysate, plasma, secretions, etc
d. How is the output signal detected?
The LC signal is detected via serial elution which is detected by a detector The MS signal is detected in the sensor part of the MS apparatus
e. What limitations does the tool have?
- High cost and complexity of the machine
- Isobaric and isometric compounds would yield the same result and is indistinguishable without further analysis
- Compound-specific ionization limits such as in nonpolar or volatile compound
3. For two of the mass spec instruments that you saw in the lab:
Instrument 1 — Xevo QTOFWhat makes it unique. It pairs a quadrupole with a TOF analyser, so a single m/z can be isolated and fragmented for deeper analysis rather than just measured. That combination gives it high sensitivity for both characterisation and quantitation of trace analytes, and makes it the instrument of choice for probing primary structure and post-translational modifications of therapeutic proteins.
What the output looks like. Mass spectra (m/z vs. intensity). With the quadrupole selecting a precursor, you additionally get MS/MS fragment spectra for that precursor; coupled to LC, you can plot individual ions over time as extracted ion chromatograms.
What I would measure. Site-specific N-glycosylation of recombinant glucocerebrosidase expressed in my non-model yeast host. GCase has four N-glycosylation sequons, and occupancy at Asn19 is required for catalytic activity — so the question “is the enzyme glycosylated?” is not enough; I need to know which site. Tryptic digestion followed by LC-MS/MS lets me identify each glycopeptide by its peptide backbone and read the attached glycan mass, with oxonium ions in the fragment spectrum confirming a glycan is present.
What it would help me understand. Two things at once. First, whether Asn19 is occupied — which separates “the host expressed inactive protein” from “the host expressed no protein”, a distinction invisible to an activity assay alone. Second, glycan composition: high-mannose structures would support mannose-receptor uptake, whereas hypermannosylated species with extended outer chains would be a liability. This is the measurement that would tell me whether the host is viable for the product at all.
Sample format and prep. Denatured and digested. Reduce with DTT or TCEP, alkylate with iodoacetamide, digest with trypsin, then desalt on C18 before injection. One prep decision matters a lot here: I would not treat with PNGase F if occupancy is the question, since removing the glycans destroys the information I’m after. The stronger design is paired injections, ±PNGase F — the deglycosylated arm confirms peptide identity and site assignment, the intact arm carries the glycan.
Instrument 2 — Cyclic IMS
What makes it unique. It’s a QTOF with ion mobility integrated into the flight path, so ions are separated by shape and size as well as mass. That adds a third dimension to the data and makes it suited to probing secondary and tertiary structure of proteins, which no other instrument on the tour does. It also accepts the widest range of sample introduction — LC, direct injection, or DESI for 2D spatial imaging.
What the output looks like. The same data can be viewed as a mobilogram (in mobility space), a chromatogram, or a mass spectrum. Drift times convert to collision cross-section values. If DESI imaging is used, spectra are recorded as a function of position, producing an image of ion intensities.
What I would measure. A native-state conformational comparison between GCase from my host and a commercial imiglucerase reference, run in the same session. Collision cross-section distributions report whether the two populate the same conformational states, and collision-induced unfolding would compare their stability.
What it would help me understand. It separates two failure modes that look identical on an activity assay. If my enzyme is inactive and its CCS profile matches the reference, the problem is glycan structure or occupancy — a Xevo QTOF question. If the CCS profile is broadened or shifted, the protein is misfolded, and the problem is secretory capacity and folding load in the host, which is a different fix entirely (chaperone co-expression, lower induction temperature). Knowing which one I’m facing determines what I do next, so I’d run this alongside the QTOF work rather than after it.
Sample format and prep. Native — intact, folded, not reduced and not digested. Buffer-exchange into a volatile ammonium acetate buffer at near-physiological pH by gel filtration or spin concentrator, removing non-volatile salts, detergents and glycerol, which suppress ionisation and form adducts. Introduce by nanoESI direct infusion. The protein also needs to be reasonably pure and concentrated before it goes on, so this sits downstream of affinity purification.
4. Given these different measurement technologies, describe how you would use at least one instrument to measure something for your final project. Describe how you would prepare the sample, generate the data, what the data would tell you, and how it would advance your project.
Instrument: CDMS (Charge Detection Mass Spectrometer)
What I want to measure. If I express GCase inside liposomes using encapsulated cell-free machinery, the question that decides whether the approach works is not “did protein get made” but “how much protein is inside each vesicle, and how uniform is that across the population.” Bulk assays cannot answer this. A Western blot on lysed liposomes gives total protein across the whole sample; an activity assay gives total activity. Neither distinguishes a population where every vesicle carries a moderate payload from one where 10% are loaded and 90% are empty — and those two scenarios have completely different implications for a delivery vehicle. CDMS answers it because it measures the mass of individual particles rather than an ensemble average.
Sample preparation. Native format throughout — the point is to measure intact, loaded vesicles, so nothing is denatured, reduced or digested. After the cell-free reaction, I would remove unencapsulated protein and residual IVTT components by size-exclusion chromatography or density-gradient centrifugation, since external protein would otherwise be indistinguishable from payload. Buffer-exchange into volatile ammonium acetate at near-physiological pH, because non-volatile salts, glycerol and detergents suppress ionisation and form adducts that broaden the mass distribution. Introduce by nanoESI. I would run three arms in the same session: empty liposomes (no DNA template), liposomes with the GCase template, and a no-liposome cell-free reaction, so that any mass shift can be attributed to payload rather than to vesicle-to-vesicle variation.
Data generated and what it looks like. CDMS output is presented both as a conventional mass spectrum and as a true mass versus charge plot. For my sample the informative view is the mass distribution across thousands of individually measured particles: empty vesicles form one population, loaded vesicles a second at higher mass. The offset between the two distributions, divided by the mass of GCase (~60 kDa), estimates copies per vesicle; the width of the loaded distribution reports heterogeneity.
What it would tell me, and how it advances the project. Three things I cannot get elsewhere. First, encapsulation efficiency as a distribution rather than a mean — if the loaded population is broad or bimodal, the encapsulation step needs reformulating regardless of how good the bulk numbers look. Second, whether expression happened inside the vesicles at all, by comparing against the template-free arm. Third, a formulation-stage readout that transfers directly to the delivery question, since the same measurement is what’s used to characterise LNPs in pharma.