Week 7 HW: Measurement & Assays

1. For one of the analytical tools you have worked with thus far (e.g. plate readers, gel electrophoresis), describe:

Analytical Tool: Agarose Gel Electrophoresis

a. How is the sample introduced?

The DNA sample is mixed with loading dye and pipetted into wells in an agarose gel.

b. What physiochemical property is being leveraged to study the sample with this tool?

DNA has a negative charge, allowing it to move through the gel when an electric field is applied. Smaller DNA fragments move through the gel faster than larger fragments.

c. What does this allow you to measure?

This allows me to estimate the size of DNA fragments by comparing them to a DNA ladder. I can also roughly estimate DNA concentration based on band intensity.

d. How is the output signal detected?

The DNA is stained with a fluorescent dye and visualized using UV or blue light to detect the DNA bands.

e. What limitations does the tool have?

Gel electrophoresis cannot determine the exact DNA sequence and has limited resolution when separating fragments of similar sizes. It also provides only an approximate measurement of DNA concentration.

2. For a liquid chromatography mass spectrometry (LC-MS) instrument:

a. How is the sample introduced into the LC? Into the MS?

The liquid sample is injected into the LC system, where it travels through a column to separate the compounds. The separated compounds then enter an electrospray ionization (ESI) source, where they are converted into charged droplets that evaporate to produce gas phase ions before entering the MS for analysis.

b. What physiochemical properties are being leveraged to study the sample with this tool?

LC separates molecules based on properties such as hydrophobicity, size, or charge, depending on the column used. MS separates ions based on their mass to charge ratio (m/z).

c. What does this allow you to measure?

LC-MS allows you to identify and quantify molecules in a sample based on their retention times and mass to charge ratios.

d. How is the output signal detected?

The MS detector measures the abundance of ions at different mass to charge ratios. Software then processes this information to generate mass spectra and chromatograms.

e. What limitations does the tool have?

LC-MS requires specialized equipment and can be expensive to operate. Sample preparation can be time consuming, and some molecules are difficult to ionize or separate. Complex samples can also cause interference, making accurate identification and quantification more difficult.

3. For two of the mass spec instruments that you saw in the lab:

Instruments: BioAccord & TQ Absolute

a. What makes the instrument unique (or, what does it specialize in)?

BioAccord: The BioAccord is designed for routine protein characterization and identification. It uses a Time of Flight (TOF) analyzer and is relatively easy to operate compared to more advanced mass spectrometers.

TQ Absolute: The TQ Absolute specializes in highly sensitive quantification of specific molecules, even at very low concentrations. Unlike other mass spectrometers, it requires knowing which molecules you want to measure beforehand.

b. What does the output data look like?

BioAccord: The output consists of mass spectra showing mass to charge ratio (m/z) versus intensity. It can also generate extracted ion chromatograms showing ion intensity over time.

TQ Absolute: The output consists of ion chromatograms showing the signal intensity of selected ions over time. The peak areas can be used to determine the concentration of a target molecule.

c. Describe something you could measure for this class (drawing from previous labs or your project) using this instrument. What would the measurement help you understand?

BioAccord: I could use the BioAccord to confirm the expression of the Gerbera hybrida DFR enzyme in my genetically engineered petunias. Identifying peptides from this protein would help me determine whether the introduced gene is producing the intended enzyme.

TQ Absolute: I could use the TQ Absolute to measure the amount of pelargonidin produced in my engineered petunias. This would help me determine whether introducing the DFR enzyme increases orange pigment production compared to unmodified plants.

d. What format would you need your sample in to perform this analysis (denatured, native, digested)? What other sample preparation might you need to do?

BioAccord: I would need to extract the proteins from the petunia tissue and digest them into smaller peptides using trypsin. The sample would also need to be cleaned up to remove contaminants before LC-MS analysis.

TQ Absolute: I would need to extract the pigments from the petunia petals using a solvent such as acidified methanol. The extract would then be centrifuged and filtered before analysis. Since I would be measuring small molecules rather than proteins, the sample would not need to be denatured or digested.

4. Given these different measurement technologies, describe how you would use at least one instrument to measure something for your final project. Describe how you would prepare the sample, generate the data, what the data would tell you, and how it would advance your project.

For my final project, I would use the TQ Absolute to measure the amount of pelargonidin produced in my genetically engineered petunias. I would collect flower petals from both modified and unmodified plants and extract the pigments using acidified methanol. After centrifuging and filtering the samples, I would analyze them using LC-MS to quantify pelargonidin, using a reference standard for comparison. This would allow me to determine whether introducing the Gerbera hybrida DFR enzyme increases pelargonidin production compared to the control plants. The results would help me understand whether my genetic modifications are successfully redirecting the flavonoid pathway toward producing orange flowers.