Week 7 HW: Measurement & Assays

Lab Section: Overview and Background The lab section of the session will feature a tour to review 5 different mass spectrometry instruments with 5 different capabilities. Students will learn how samples are introduced for each instrument (e.g. infusion, DESI, LC), how different instrument sensitivities impact the type of analysis performed, and what the data looks like for each instrument.

BioAccord – Protein sequence, protein mass The BioAccord is designed for non-LC-MS users and features an LC for sample introduction and separation, self-diagnostic capabilities, and low-mid performance in terms of sensitivity. It is ideal for routine characterization problems (e.g. protein identification) and non-trace quantitative analyses and is commonly used for process analytical analyses in pharma. It uses a Time-of-Flight (TOF) analyzer that generates mass spectra (m/z vs. intensity plots). Individual ions can be plotted over time, which is referred to as an extracted ion chromatogram.

TQ Absolute - Protein/peptide quantitation The TQ absolute is designed for robust quantitation of trace analytes. It requires the user first know the analytes they are looking for but features extremely high performance for thousands of injections. It is often used for clinical samples (lab blood testing) and the quantification of drugs and drugs metabolites during the DMPK phase of pharma discovery. The TQ of the TQ Absolute stands for “Tandem Quad,” as the instrument features two quadrupoles that operate by selecting only ions of particular mass-to-charge ratios. Because of this, data is not recorded as a spectrum but as an ion trace over time, called an ion chromatogram.

Xevo QTOF – Post-translational modifications, proteoforms, quantitation The Xevo QTOF is a high sensitivity instrument used for both characterization and quantitation of trace analytes. It is ideal for probing the primary structure and PTMs of most therapeutics and performing quantitation. The system features a TOF detector but also a device called a quadrupole (Q), which allows for a single m/z to be selected for deeper analysis. The QTOF generates mass spectra.

Cyclic IMS – Tertiary structure The Cyclic IMS is a unique mass spectrometry system (QTOF) into which ion mobility is integrated into the flight path. This offers a unique tool set for probing the secondary and tertiary structure of proteins, lipids, and other molecules. It is ideal for deep characterization and across a range of applications; because of this, sample is often introduced by a variety of different methods, including LC, direct injection, or by 2D spatial imaging methods such as DESI (Desorption ElectroSpray Ionization). The addition of the ion mobility device adds a third dimension to the data and data can be viewed as a mobilogram (viewed in mobility space), a chromatogram (similar to BioAccord or Xevo, or as a mass spectrum. If an imaging technique is used, mass-to-charge spectra are recorded as a function of position, generating an image of ion intensities. This technique can be used to image drug and drug metabolite abundance in different tissues (e.g. liver). CDMS – viruses, large multi-protein complexes, glycoproteins

The CDMS (Charge Detection Mass Spectrometer) is a special instrument used for the analysis of very large (megadalton) biomolecules in the folded (native state). It uses a unique detector that samples one ion at a time, allowing these very large ions to be properly measured. It is often used for AAV or LNP characterization in the pharmaceutical space. Data is often presented both as a traditional mass spectrum (m/z vs. intensity) and as a true mass vs. charge plot.

Homework questions Introduction Goal: For this homework you will reflect on the different analytical tools you have used in class so far and integrate the learnings from the analytical module. Analytical technologies are designed to measure different aspects of molecules and samples. All analyzers include the following components: 1) a mechanism to introduce the sample, 2) a mechanism to probe the sample, leveraging physiochemical properties of the target analytes, 3) a mechanism to detect the probed response (i.e. a detector/digitizer), and 4) a mechanism to analyze the data. In this module, liquid chromatography and mass spectrometry were introduced as advanced tools for quantifying and characterizing the structure of large and small molecules. Mass spectrometers serve to separate and detect ions by their mass-to-charge (m/z) ratios, allowing identification by molecular weight. Virtually every analytical technique performs better with improved purity of sample; liquid chromatography can facilitate the specific detection of molecules by separating them in the time domain (prior to UV or MS detection) using properties such as the differential hydrophobicity, isoelectric point, or size of the molecules.

For one of the analytical tools you have worked with thus far (e.g. plate readers, gel electrophoresis), describe: The analytical tool I will describe is the spectrophotometer. How is the sample introduced? The sample is introduced by placing it in an absorption cell or cuvette (round or rectangular, made of glass, quartz, fused silica, or plastic), which is positioned in the path of the instrument’s light beam. If the compound to be analyzed does not absorb light on its own, it is first mixed with other reagents to obtain a solution that does (1) (García, 2018). What physiochemical property is being leveraged to study the sample with this tool? It utilizes the absorption and transmission of electromagnetic radiation (light) at specific wavelengths, a property that depends on the molecular structure of the sample (specifically its functional groups). It is mathematically based on the Beer-Lambert Law (A=lc), relating absorbance to concentration (García, 2018).

What does this allow you to measure? It allows the measurement of absorbance (the amount of light absorbed) or transmittance (the percentage of light that passes through the sample). Based on this, it enables the determination of the concentration of specific compounds (e.g., iron in blood or pollutants in water), identification of structural units, color standardization, and measurement of the full absorption spectrum (García, 2018).

How is the output signal detected? The light that passes through the sample and is not absorbed reaches a photodetector (such as vacuum tube photocells in historical models or modern optical detectors), which registers the transmittance and converts the light signal to display the result on a digital screen (García, 2018).

What limitations does the tool have? Cuvette material interference: Common optical glass absorbs light in the ultraviolet range (below 350 nm), making it necessary to use more expensive quartz cuvettes for UV readings. Absorbance requirement: If a substance is transparent or does not absorb radiation at the instrument’s wavelengths, it requires prior treatments or chemical reactions with specific reagents. Stray light and resolution loss: Simpler or older models may exhibit loss of light intensity or lower resolution depending on the light source and optical selector used. (García, 2018)

For a liquid chromatography mass spectrometry (LC-MS) instrument: How is the sample introduced into the LC? Into the MS? The sample is first placed in small vials and pumped through the LC system into a chromatographic column, where its components are separated. The liquid leaving the column is transferred through a capillary to the mass spectrometer’s ionization source. In the instrument demonstrated during the visit, the sample was sprayed as a fine mist toward a metal cone, consistent with electrospray ionization (ESI), which generates gas-phase ions for mass analysis. What physiochemical properties are being leveraged to study the sample with this tool? LC-MS combines two types of physicochemical separation and analysis. Liquid chromatography separates compounds based on their different interactions with the mobile and stationary phases, which may depend on properties such as polarity, hydrophobicity, and charge. Mass spectrometry then analyzes ionized molecules according to their mass-to-charge ratio (m/z).

What does this allow you to measure? LC-MS allows researchers to separate, identify, characterize, and, with an appropriate analytical method, quantify compounds in complex mixtures. It can be used to analyze molecules such as proteins, peptides, metabolites, and other compounds based on their chromatographic behavior and mass-to-charge ratios.

How is the output signal detected? The mass spectrometer detects ions after they pass through the mass analyzer. The detector converts the incoming ions into an electrical signal, which is processed to generate a mass spectrum showing signal intensity as a function of mass-to-charge ratio (m/z). When coupled to LC, the data can also be displayed as a chromatogram, showing signal intensity over time.

What limitations does the tool have? LC-MS has limitations related to sample preparation, sample complexity, and the analytical method. Components in complex mixtures may interfere with one another, including through ion suppression. In addition, instrument selection involves trade-offs. A triple quadrupole provides highly sensitive targeted analysis but only measures selected analytes, whereas a time-of-flight analyzer provides broader, untargeted information but may have lower sensitivity. For two of the mass spec instruments that you saw in the lab:

What makes the instrument unique (or, what does it specialize in)? The triple quadrupole mass spectrometer specializes in targeted analysis and highly sensitive quantification. It uses three quadrupole sections to select and analyze specific ions of interest. This allows researchers to focus on selected analytes even when other compounds are present in the sample. During the lab visit, the instrument was described as having approximately 10–100 times greater sensitivity than several of the other mass analyzers demonstrated. The time-of-flight (TOF) mass spectrometer specializes in broad, untargeted analysis. Unlike a triple quadrupole, it does not focus only on previously selected analytes. Instead, it measures a broader range of ions. It separates ions according to their time of flight through the analyzer, which depends on their mass-to-charge ratio. The instrument demonstrated used a folded flight path with reflections to increase the travel distance and improve ion separation.

What does the output data look like? The triple quadrupole mass spectrometer output consists of chromatograms or signal panels corresponding to selected mass-to-charge ratios. During the demonstration, different measurements targeted m/z values such as 195, 281, 393, and 255. Each panel showed the signal associated with the selected analyte, while other compounds in the sample were excluded from that targeted measurement. The time-of-flight (TOF) mass spectrometer output is a mass spectrum that can show a charge-state envelope. During the demonstration, an intact but denatured protein was analyzed without enzymatic digestion. The spectrum contained multiple peaks corresponding to different charge states of the same protein, rather than different proteins. This pattern can be used to characterize the protein and estimate its molecular mass.

Describe something you could measure for this class (drawing from previous labs or your project) using this instrument. What would the measurement help you understand? For my project, a triple quadrupole mass spectrometer could be used to quantify a specific peptide derived from the reporter protein produced by the cell-free protein synthesis (CFPS) system. After enzymatic digestion of the reporter protein, a targeted LC-MS/MS method could measure the selected peptide. This would help determine whether the CFPS system is producing the expected reporter protein and compare its production across different alginate formulations. For my project, a TOF mass spectrometer could be used to characterize the intact reporter protein produced by the CFPS system. By measuring its molecular mass, we could verify whether the expected protein was produced and investigate possible variations in its molecular composition. This would provide an additional way to validate the biological component responsible for generating the colorimetric signal.

What format would you need your sample in to perform this analysis (denatured, native, digested)? What other sample preparation might you need to do? In the triple quadrupole mass spectrometer, the sample would be prepared as digested peptides rather than as an intact protein. The reporter protein would first be recovered from the CFPS formulation and separated from interfering components. It would then undergo enzymatic digestion to generate peptides. The resulting sample would be prepared in an LC-MS-compatible solution for targeted analysis.

In the TOF mass spectrometer, the instrument demonstrated was used to analyze an intact but denatured protein, without enzymatic digestion. For my project, the reporter protein would first be recovered from the CFPS reaction and separated from interfering components. It would then be prepared under conditions compatible with the instrument. Following the demonstration, the protein could be analyzed in a denatured, unfolded state. Given these different measurement technologies, describe how you would use at least one instrument to measure something for your final project. Describe how you would prepare the sample, generate the data, what the data would tell you, and how it would advance your project.

For my final project, I would use a triple quadrupole mass spectrometer to quantify the reporter protein produced by the cell-free protein synthesis (CFPS) system incorporated into the biodegradable breast pad. First, I would prepare CFPS reactions containing the reporter system and recover the produced protein. The protein would be separated from interfering components and enzymatically digested into peptides. A specific peptide derived from the reporter protein would then be selected as the target for a targeted LC-MS/MS analysis. The samples would be introduced into the liquid chromatography system, where their components would be separated before entering the mass spectrometer. The triple quadrupole would detect the selected peptide based on its mass-to-charge ratio. By using appropriate reference standards and a calibration curve, I could estimate the amount of reporter protein produced in different CFPS formulations. Comparing these results would help identify which formulation supports better reporter protein production. This measurement would help validate the biological component of the biosensor before integrating it into the alginate-based breast pad. However, protein quantification alone would not demonstrate that the biosensor responds correctly to the target stimulus. It would need to be complemented by functional testing of the colorimetric response.