Week 7: Mass Spectronomy

Week 7 HW: DNA Design

HTBAA Homework #7 — Mass Spectrometry (Waters)

Matthew Woslum · September 29, 2026 Homework 7

Question 1: Gel Electrophoresis

Since I am a remote committed listener in HTBAA, I will use analytical tools from my own lab as examples.

  1. How is the sample introduced: A mix of purified DNA and loading dye is pipetted into a well.
  2. What physiochemical property is being leveraged to study the sample with this tool: The size of the DNA fragments and the UV absorbance of the DNA stain.
  3. What does this allow you to measure: The size of the DNA fragments that have been cut at precise loci beforehand with restriction enzymes.
  4. What limitations does the tool have: Similar sized fragments are indistinguishable and low resolution.
     

Question 2: LC-MS

  1. How is the sample introduced into the LC and MS: The samples are injected first into the LC section. The LC separates the samples first by hydrophobicity, then the MS separates them by mass-to-charge ratio (m/z).
  2. What physiochemical properties are being leveraged to study the sample with this tool: Hydrophobicity and mass-to-charge ratio.
  3. What does this allow you to measure: This tells you the molecular weight of the components and lets you quantify them from complex samples.
  4. How is the output signal detected: After first separating molecules with LC (Liquid Chromatography), they are injected into the MS (Mass Spectrometer), which ionizes them and separates them by their mass-to-charge ratio. Output is displayed as a Ion Chromatogram.
  5. What limitations does the tool have: Limitations include non-trace sensitivity, and problems with huge molecules.
     

Question 3: Two of the mass spec instruments seen in lab

  1. What makes the instrument unique (or, what does it specialize in): TQ Absolute: It uses a Tandem Quadrupole (TQ) setup that filters for pre-selected ions. It also has high sensitivity for targeted trace samples. Cyclic IMS: Combines a QTOF with cyclic ion mobility separation (IMS) to measure mass and 3D molecular shape (secondary and tertiary structure). It also supports 2D spatial tissue imaging via DESI.
  2. What does the output data look like: TQ Absolute: Single-ion tracking over time (ion chromatograms) for pre-selected ions rather than full mass spectra. Cyclic IMS: 3D data combining mass-to-charge (m/z), drift time (mobilograms), and intensity, or heat maps (mass spectrometry imaging).
  3. Describe something you could measure for this class (drawing from previous labs or your project) using this instrument. What would the measurement help you understand: TQ Absolute: I could measure the exact concentration of a specific target protein or a specific antibiotic/product produced by engineered cells from a complex cell lysate or culture media after miniprep. Cyclic IMS: I could measure the folded 3D (secondary/tertiary) structure of a purified protein that was expressed.
  4. What format would you need your sample in to perform this analysis (denatured, native, digested)? What other sample preparation might you need to do: TQ Absolute: Protein would need to be digested. Enzymatically cleaved into smaller signature peptides because the Tandem Quadrupole targets specific, low-m/z peptide fragment transitions. Cyclic IMS: Intact and folded so that the protein retains its natural 3D secondary and tertiary structure during gas-phase ion mobility separation.
     

Question 4: Using an MS instrument for final project

For my final project on 3D-bioprinting bone tissue from adipose-derived stem cells (ADSCs), I would use the TQ Absolute to check if the stem cells are actually turning into bone cells and producing real bone matrix.
Sample Prep: I would take samples of the printed scaffolds at different days during growth (Days 0, 7, 14, and 21). After breaking open the cells to extract the proteins, I would use trypsin to chop the proteins into smaller peptides, clean up the sample, and add heavy-labeled standard peptides for key bone markers like Collagen Type I and Osteocalcin.
Data: The liquid chromatography (LC) separates the peptides by hydrophobicity, and then the TQ Absolute filters for and measures the exact mass of the specific bone peptides. This gives me clean chromatogram peaks showing how much of each bone protein is present.
How It Helps My Project: Tracking how the amounts of early and late bone markers increase over time would prove that my bioprinting setup and media were actually making real bone tissue rather than just random mineral buildup. This data would help me tweak my bioink recipe and scaffold design to get stronger, more realistic bone growth.