Week 7 HW: Checking Proteins and Cortisol
HTBAA Homework 7 — Waters
My final project is to make two proteins for cortisol detection and later use them in a microfluidic device. Cortisol helps the two protein partners come together. With the right substrate, the assembled system produces light. The published design used E. coli BL21(DE3) to make the proteins, so I would use this as my starting production host. [1]
| Protein | Parts and role |
|---|---|
| corD1-3.5GS-αLux | The corD1 partner joined to one part of the light-producing enzyme. |
| βitsPro-4GS-mhcy129 | The cortisol-binding part mhcy129 joined to the other part of the light-producing enzyme. |
The cells would make the proteins during manufacturing. The sensing device would use the purified proteins, fresh samples, and fresh reagents for each measurement. My main questions are: Did I make the expected proteins? Do they work? How well do they measure cortisol?
1. A tool from earlier coursework
Assignment
For one of the analytical tools you have worked with thus far (e.g. plate readers, gel electrophoresis), describe:
My choice: a microplate reader in fluorescence mode. In Homework 4, I analyzed the course’s sfGFP reference data for the cell-free protein synthesis assignment. The report says a Spark plate reader was used to measure fluorescence. My answer is based on analyzing these reference data. [2]
a. How is the sample introduced?
We put small amounts of liquid into the wells of a plate. Each well holds one sample. We then place the plate inside the reader. In the course’s deep-well experiment, samples were moved from the 96-well incubation plates into a 384-well reading plate before measurement. [2]
b. What physiochemical property is being leveraged to study the sample with this tool?
The reader uses fluorescence. When light shines on sfGFP, the protein absorbs some of that light and gives off green light. The reader measures this green light.
c. What does this allow you to measure?
We directly measure how strong the green light is. We compare it with samples containing known amounts of sfGFP. This helps us estimate the concentration of fluorescent sfGFP and compare different reaction conditions.
d. How is the output signal detected?
The reader shines light into each well and collects the green light from the sample. Its optical system separates this light from the excitation light. A detector turns the collected light into an electrical signal. The computer shows the result as a number. [3]
e. What limitations does the tool have?
A stronger signal does not always mean more protein. The protein needs to fold and develop its fluorescent part before it can give off green light. Other substances in the sample and the reader settings can also affect the result.
We need blanks and samples with known concentrations for comparison. Very strong samples may need dilution. This tool also cannot tell us the protein’s exact mass or sequence.
In the course report, the first sample signals were above the highest standard, so the samples were diluted and read again. This is a useful example of checking the measurement range instead of trusting every number from the reader. [2]
2. Liquid chromatography–mass spectrometry
Assignment
For a liquid chromatography mass spectrometry (LC-MS) instrument:
LC separates the mixture before MS analyzes the ions. An ion is a molecule or atom carrying an electrical charge.
a. How is the sample introduced into the LC? Into the MS?
A small amount of liquid sample is injected into the LC. Flowing liquid carries it through a column, where different substances separate.
The liquid leaving the column then enters an ion source. One common method is electrospray. It makes tiny charged droplets. As the liquid evaporates, charged molecules enter the MS.
b. What physiochemical properties are being leveraged to study the sample with this tool?
LC uses differences in how molecules interact with the column material and the flowing liquid. Some stay in the column longer, so they come out later. For example, a common type of LC uses differences in how strongly molecules are attracted to a water-repelling surface. [4]
MS separates or measures ions by their mass-to-charge ratio, called m/z. This means the mass of an ion divided by its number of charges. [5]
c. What does this allow you to measure?
LC-MS helps us find out which substances are in a sample. We can estimate their molecular masses and compare them with what we expect.
With a suitable method and known standards, we can also measure how much of a target substance is present. For my project, it could help check the proteins I make or measure cortisol in a sample.
d. How is the output signal detected?
After the mass analyzer separates the ions, a detector records them and turns their signals into electrical signals.
The computer makes graphs showing the signal at different m/z values and how the signal changes over time.
e. What limitations does the tool have?
The sample needs careful preparation. Too much salt or other unwanted material can weaken the signal. Different molecules can also produce different signal strengths, so signal strength alone does not tell us concentration.
Some substances have the same mass and need more information to tell them apart. LC-MS also cannot prove by itself that a protein works. We still need a separate test of its function.
3. Two instruments from the lab visit
Assignment
For two of the mass spec instruments that you saw in the lab:
I chose the Xevo QTOF and TQ Absolute. Their abilities overlap: QTOF can also measure cortisol, and TQ instruments can measure proteins through selected peptides. I chose different tasks to match their strengths. This does not mean my project must use both machines. A single suitable instrument could do both tasks with different sample preparation and measurement methods. [6,7]
Instrument 1 — Xevo QTOF
a. What makes the instrument unique (or, what does it specialize in)?
QTOF means quadrupole time-of-flight. The quadrupole can select ions, and the time-of-flight part uses their travel times to measure m/z.
The Xevo QTOF measures m/z with high resolution. This helps separate ions with very similar m/z values. It can also break selected ions into smaller pieces and measure those pieces. This gives us more information about a molecule. [5,6]
b. What does the output data look like?
It produces a mass spectrum, with m/z on the horizontal axis and signal strength on the vertical axis. Software can also use the data to estimate a protein’s mass. If we break molecules into smaller pieces, we get a spectrum of those pieces too.
c. Describe something you could measure for this class (drawing from previous labs or your project) using this instrument. What would the measurement help you understand?
I would measure each of the two fusion proteins for my cortisol sensor. I would compare their measured masses with the masses expected from their sequences, including any tags kept on the final proteins.
This would help me check whether I made the expected proteins. If needed, I would also study smaller protein pieces for more evidence of their identity. I would still need a separate light-producing test to check whether the proteins work together.
d. What format would you need your sample in to perform this analysis (denatured, native, digested)? What other sample preparation might you need to do?
For the first check, I would use each purified protein separately and remove salts that could affect the measurement. I would keep the protein chain whole, although it would not need to stay folded in its natural shape. This is an intact-protein check under suitable denaturing conditions.
For a more detailed check, I would prepare another portion and use an enzyme to cut the protein into smaller pieces, called peptides. This is a digested sample. I would clean up these samples before LC-MS analysis and keep a separate portion for the functional test. [7,8]
Instrument 2 — TQ Absolute
a. What makes the instrument unique (or, what does it specialize in)?
TQ means tandem quadrupole. The TQ Absolute belongs to the triple-quadrupole family. It is especially useful for measuring small amounts of a known target in a mixed sample.
In a common measurement mode called MRM, it selects a target ion, breaks it into smaller pieces, and selects a specific piece to detect. These two selection steps help reduce signals from other substances. [9]
b. What does the output data look like?
In MRM mode, it produces a chromatogram. The horizontal axis shows time, and the vertical axis shows signal strength. The target appears as a peak.
We measure the area under the peak and compare it with known standards to calculate the target concentration.
c. Describe something you could measure for this class (drawing from previous labs or your project) using this instrument. What would the measurement help you understand?
I would use it to measure cortisol concentration in samples for my sensor project, after checking that the method works for those samples. I would start with artificial sweat and later consider collected sweat.
I would split the same sample into two portions. One would be tested with my light-producing protein sensor. The other would be tested using LC-MS/MS with the TQ Absolute. Here, LC separates the sample, and the TQ Absolute is the mass spectrometer. Comparing the results would help me check how well my sensor estimates cortisol concentration.
LC-MS/MS would provide an independent reference measurement, rather than an error-free true value. A published study measured sweat cortisol using a triple-quadrupole system. That supports this approach, but I would still need a suitable method on the available TQ Absolute. [10]
d. What format would you need your sample in to perform this analysis (denatured, native, digested)? What other sample preparation might you need to do?
Cortisol is a small molecule, so it does not need the protein preparation steps of unfolding or enzyme digestion.
I would collect a separate sample before adding the sensing proteins or light-producing substrate. I would use a suitable method to extract cortisol and remove unwanted material. I would also prepare known cortisol standards and use an internal standard—a reference substance added to help correct measurement errors. [10]
4. Using measurement tools for my final project
Assignment
Given these different measurement technologies, describe how you would use at least one instrument to measure something for your final project. Describe how you would prepare the sample, generate the data, what the data would tell you, and how it would advance your project.
My answer
My project is to make two proteins that produce light in response to cortisol and later use them in a microfluidic device. I would use different tools to check the proteins and test the measurement.
First, I would check the proteins I make
I would produce the two proteins in E. coli and purify them separately. I would prepare a small portion of each protein for Xevo QTOF analysis, including removing salts that could affect the signal.
I would compare the measured masses with the masses expected from the protein sequences. If needed, I would also analyze smaller protein pieces to help confirm their identity. This would help me check my products before testing the sensor.
Next, I would test whether the proteins respond to cortisol
I would mix the two proteins with the light-producing substrate and samples containing known cortisol concentrations. I would include a sample with no cortisol and keep the reaction and reading times the same.
A plate reader would measure the light. I would use one set of known samples to build a calibration curve. I would then test separate samples to check how well the curve estimates their concentrations.
I would also compare the sensor with LC-MS/MS using the TQ Absolute
I would begin with artificial sweat containing known amounts of cortisol. Later, I would compare samples with unknown concentrations.
Before adding the sensing proteins or substrate, I would split each sample into two portions. One would go to my protein sensor. For the other, I would add an internal standard, extract cortisol, and prepare it for LC-MS/MS using a checked method on the TQ Absolute.
I would compare the cortisol concentrations from the two methods, taking any dilution into account. I would use repeated measurements across several concentrations. The comparison would use matched samples and a clear definition of the cortisol being measured, since extraction and protein binding can affect what each method measures.
These results would guide the next stage of my project
The protein measurements would help check that I made the expected products. The light measurements would test their sensing function. The LC-MS/MS comparison would help check the concentration estimates.
If the results were consistent, I would move on to the microfluidic device. If they differed, I would investigate the protein reagents, sample conditions, calibration, or measurement timing before combining the parts.
Connection to the four project questions
| Project question | How this measurement plan helps me answer it |
|---|---|
| What do I want to make, and why? | Two functional protein reagents for cortisol detection, later connected to microfluidic sampling. Mass checks and light measurements test different parts of that goal. |
| What living system could make it? | I would start with the published E. coli BL21(DE3) route. I would judge production by the recovered proteins and their sensing function, not only cell growth. |
| How could automation help? | A liquid handler could prepare standards and test mixtures. Timed plate reading could make comparisons more consistent. Later, microfluidics could control sample and reagent delivery. |
| What would it take at a meaningful scale? | I need to measure the amount of each protein used per successful test, usable protein yield, losses, and batch time. These measurements would support a production estimate. |
For each protein, I would calculate total demand = number of tests × protein used per test + protein for calibration, controls, and losses. I would then use the measured usable yield to estimate culture volume. Production time would include growth and expression, purification, and quality and function checks. I would not use faster cell growth alone as evidence of a better manufacturing process.
AI use
I used ChatGPT/OpenAI Codex to help me understand the instruments. We discussed why QTOF and TQ have overlapping abilities and how to choose measurements for my project.
References
- Chen, J. Y.-H., et al. (2025). De Novo Design of High-Performance Cortisol Luminescent Biosensors. Journal of the American Chemical Society. DOI · Author-hosted paper.
- Course reference report: sfGFP process optimization, quad 96-well incubation: processopt_20261008 (October 9, 2026), especially pp. 2–3. Report from Homework 4 · My Homework 4 analysis.
- Tecan. Spark Multimode Microplate Reader. Instrument overview. Available measurement modes depend on the instrument configuration.
- Waters. HPLC Separation Modes. Beginner’s guide to liquid chromatography.
- Waters. What Instruments and Tools Are Commonly Used in Mass Spectrometry? Mass spectrometry primer.
- Waters. Xevo G3 QTof. Instrument overview. Used here to explain QTOF capabilities; the exact model and setup available in the lab determine the method.
- Dunning, C., and Wrona, M. D. (2020). Developing a Quantitative Surrogate Peptide Assay—Peptide Mapping through MRM Optimization for Measuring Dulaglutide in a Rat PK Study. Waters application note. A fusion-protein example, not a validated method for my two proteins.
- Waters. UPLC Intact Mass Analysis Kit with Tubing & Manual, SKU 176001519. Official product description explaining protein desalting before LC-MS.
- Waters. Xevo TQ Absolute Overview and Maintenance Guide, version 02. Official manual, especially the MS/MS operating modes.
- Jia, M., et al. (2016). Quantification of Cortisol in Human Eccrine Sweat by Liquid Chromatography–Tandem Mass Spectrometry. Analyst 141, 2053–2060. Full text · DOI. This study used a different triple-quadrupole platform, so transfer to TQ Absolute would require method checks.