Week 7 HW: 📏 Measurement & Assays

Week 7 HW: 📏 Measurement & Assays

1️⃣ Question 1: SDS-PAGE 🧫

I chose SDS-PAGE, a technique I learned about in one of my university courses to check the GFP made by transformed bacteria.

  • a) 💉 Sample introduction: I lyse the cells, mix the protein with sample buffer (SDS + DTT/β-mercaptoethanol + glycerol + dye), heat it to ~95 °C, and pipette it into the wells of a polyacrylamide gel. One well gets a protein ladder.
  • b) 🔬 Property used: Size. SDS unfolds proteins and gives them a charge proportional to their length, so only molecular weight matters. In an electric field, small proteins move through the gel mesh faster than big ones.
  • c) 📊 What I can measure: approximate molecular weight, whether my protein is expressed (a band at ~27 kDa), purity (extra bands), rough relative amount (band intensity), and disulfide-linked chains (reducing vs non-reducing).
  • d) 🔦 Detection: I stain the gel (Coomassie, or silver for more sensitivity), image it, and plot log(MW) vs distance migrated for the ladder to read off my band’s size. Band intensity is quantified with ImageJ.
  • e) ⚠️ Limitations: low resolution, only semi-quantitative, no proof of protein identity, denaturing (function is lost), low throughput, and much less sensitive than mass spec.

2️⃣ Question 2: LC-MS 🧪

  • a) 💉 Sample introduction:
    • Into the LC: the sample is dissolved in a mobile-phase-compatible solvent, centrifuged or filtered, and injected by an autosampler. A high-pressure pump pushes it through a column.
    • Into the MS: the column eluent goes into an electrospray ionization (ESI) source. A high voltage makes charged droplets, the solvent evaporates, and gas-phase ions enter the vacuum of the instrument.
  • b) 🔬 Properties used: LC uses hydrophobicity (reversed-phase), or size (SEC) or charge/pI (ion exchange) in other modes. MS uses the mass-to-charge ratio (m/z).
  • c) 📊 What can be measured: intact mass, retention time, purity, sequence and PTMs (via MS/MS or digests), and quantity from the peak area.
  • d) 🔦 Detection: ions hit a detector that converts each hit into an electrical pulse, and a digitizer records it. In a TOF, flight time is converted to m/z. The output is a mass spectrum plus a chromatogram, and software deconvolutes the charge states into the neutral mass.
  • e) ⚠️ Limitations: quantitation needs standards, it can’t tell isomers apart by mass alone, and it is expensive and needs training.

3️⃣ Question 3: BioAccord (A) vs CDMS (E) ⚖️

🅰️ BioAccord🅴 CDMS (Charge Detection MS)
a) ✨ What’s uniqueA plug-and-play LC-TOF with built-in LC and self-diagnostics, made for routine protein identification and mass (low–mid sensitivity)Measures one ion at a time, getting both m/z and charge (z), so it weighs megadalton, native particles directly (AAV, LNPs, viruses, large complexes)
b) 📈 OutputMass spectrum (m/z vs intensity) with a charge-state envelope, TIC/XIC chromatograms (intensity vs time), and a deconvoluted neutral massAn m/z spectrum plus a true mass vs charge plot (one dot per ion), summed into a mass histogram of the species present
c) 🧪 What I’d measureAptamer product: exact mass of the 39-nt aptamer unit after RCA and XcmI cutting, plus shorter/longer strands (n−1/n+1). I could also check insulin (~5808 Da) for +1 Da deamidation, +16 Da oxidation or dimers after a temperature excursionRCA product: the mass distribution of the very long concatemers (average repeats per strand, and batch consistency). I could also look for heat-induced insulin aggregates
d) 🧫 Sample formatDenatured (acidic/organic mobile phase). Can be reduced (insulin A and B chains) or digested (to locate modifications). Oligos use an ion-pairing mobile phase. Needs filtered, desalted sample with no detergents or non-volatile salts, so I’d remove phi29, Mg²⁺ and dNTPsNative/folded in a volatile buffer like ammonium acetate (buffer exchange, no denaturant or digestion), very dilute so only one ion hits the detector at a time, and not filtered (aggregates are what I’m looking for)

4️⃣ Question 4: Validating my aptamer is made in the correct form 🧬✅

My project: a cell-free-made insulin-binding DNA aptamer (IGA3) for a disposable electrochemical cartridge that screens insulin after a cold-chain failure. My Week 2 design makes it in three steps:

68-nt circular template ──► phi29 RCA ──► long ssDNA concatemer ──► XcmI + adaptors ──► 39-nt aptamer unit

Each step can go wrong (RCA copies any template error into every unit), so, like a QC release test for a drug batch, I’d prove I made the right molecule, at the right length, in the right amount before scaling to 1,000 cartridges. 🎯

What “correct” looks like (my calculated average masses, assuming XcmI leaves a 5′-phosphate and 3′-OH):

🔧 Instruments: 🅴 CDMS for the huge RCA product, and 🅰️ BioAccord (main tool) for the cut products and purified aptamer.

1️⃣ Sample preparation 🧪

  • RCA product (for CDMS): remove phi29, dNTPs, primer and Mg²⁺, then buffer exchange into ammonium acetate. Dilute so single ions are measured, and handle gently so the long strands don’t shear.
  • XcmI digest (for BioAccord): take time-point samples, stop the reaction with EDTA, then desalt and remove the enzymes (salt adducts are a problem for G-rich DNA).
  • References: order a synthetic 39-nt unit and a 30-nt IGA3 standard to run side by side with my purified RCA-made aptamer.

2️⃣ Generating the data 📈

  • CDMS: a mass histogram of the RCA concatemers.
  • BioAccord: ion-pairing reversed-phase LC into the ESI-TOF (negative mode), then deconvolution to neutral masses.

3️⃣ What the data tell me 🔍

  • CDMS: average mass ÷ ~21.3 kDa gives the average number of aptamer repeats per strand (a proxy for yield). The peak width shows how uniform the reaction is, and batches can be compared.
  • BioAccord: a ~12,324 Da peak with ~9,046 Da in roughly a 1:1 ratio means both XcmI cuts worked. Uncut (~21,352 Da) or half-cut masses mean incomplete digestion, and unit ± one nucleotide means a wrong cut or copying error.
  • Synthetic standard comparison: the same retention time and mass as my purified aptamer means it is chemically identical to the reference and having one dominant peak means high purity.

4️⃣ How it advances my project 🚀

It’s my quality-control step before scale-up. It shows which step to fix (template, RCA, or XcmI cutting, by tuning digestion time and adaptor amount), confirms exactly which extra end nucleotides (AATC and TTCGT) I’m testing, and answers my Week 2 question of whether RCA-derived IGA3 matches synthetic IGA3.

⚠️ Limits: mass doesn’t prove sequence order (I’d also sequence the template/product), and it doesn’t prove folding or function (I’d follow up with an insulin-binding assay and electrode test).