Week 07 HW: Mass-spectrometry

1.For one of the analytical tools you have worked with thus far (e.g. plate readers, gel electrophoresis), describe:

How is the sample introduced?

Cultures are grown in 96- or 384-well microplates, with a liquid-handling robot dispensing the medium and cells. Nile Red stain is added to each well, often after the cells are washed and resuspended in buffer, and the plate is then loaded onto the reader’s tray. Every plate needs controls: the parent strain, a no-cell blank and a dye-only well.

What physiochemical property is being leveraged to study the sample with this tool?

Fluorescence. Nile Red is a solvatochromic dye: it fluoresces strongly in hydrophobic environments such as the neutral-lipid droplets that store triglyceride, and only weakly in water. The reader also measures absorbance or light scattering (OD) to track cell growth.

What does this allow you to measure?

Relative neutral lipid content per well, usually normalized to cell density (fluorescence divided by OD) Growth curves from the OD readings Lipid accumulation over time, for example after nitrogen starvation is triggered Rapid ranking of many strain variants in one run

How is the output signal detected?

A lamp or LED provides excitation light, which passes through a filter or monochromator into the well. The emitted fluorescence passes through a second filter to a detector, typically a photomultiplier tube or photodiode. The detector converts the light into an electrical signal, reported as relative fluorescence units (RFU). Absorbance is read the same way, using the light transmitted through the well.

What limitations does the tool have?

Relative, not absolute. RFU isn’t a mass of lipid unless calibrated against a reference method such as GC-FAME. Variable staining. Dye uptake depends on the cell wall and strain, so comparisons are safest within one strain background. Interference. Medium, pigments and cell density can add background, quench the signal or absorb light. No composition information. It can’t show the oleic acid share, or measure squalene or hydroxytyrosol. Plate effects. Evaporation at plate edges and photobleaching can distort readings.

2.For a liquid chromatography mass spectrometry (LC-MS) instrument:

How is the sample introduced into the LC? Into the MS?

Into the LC: The broth or cell extract is first cleaned up (centrifuged, filtered, diluted, and usually spiked with an internal standard) and placed in a vial in the autosampler. A small volume, typically a few microlitres, is injected into a flowing liquid (the mobile phase) pushed by a high-pressure pump through a column, often a reversed-phase C18 column.

Into the MS: The liquid leaving the column goes straight to an ionization source, usually electrospray ionization (ESI). It sprays the liquid into charged droplets, which evaporate to leave gas-phase ions. These enter the vacuum of the mass spectrometer through a small opening and are guided to the mass analyzer.

What physiochemical properties are being leveraged to study the sample with this tool?

LC: Compounds interact differently with the column and the mobile phase, depending on polarity and hydrophobicity (in reversed-phase mode). More hydrophobic compounds stay on the column longer, so each compound leaves at a characteristic retention time. MS: Ions are separated by their mass-to-charge ratio (m/z). The compound must also ionize in the source. In MS/MS, selected ions are broken into fragments, whose pattern reflects the molecule’s structure.

What does this allow you to measure?

It identifies and quantifies specific compounds in a complex mixture. For hydroxytyrosol (molecular weight about 154), that means confirming it by retention time, m/z and fragment pattern, then measuring its concentration against a calibration curve made from a pure standard. It can measure many compounds in one run, such as tyrosol alongside hydroxytyrosol.

How is the output signal detected?

Ions strike a detector, such as an electron multiplier. Orbitrap instruments instead read an image current. This converts ion arrivals into an electrical signal. The software records intensity against time (a chromatogram) and against m/z (a mass spectrum), and quantification uses the area of the peak.

What limitations does the tool have?

Needs standards. Absolute quantification requires a pure reference compound. Matrix effects. Salts, media components and other compounds can suppress or enhance the signal (ion suppression), which is a particular concern for salty marine media. Not everything ionizes. Neutral, nonpolar molecules such as triglycerides and squalene ionize poorly by ESI, which is why squalene and fatty acid profiles are better handled by GC-MS or HPLC. Isomers. Compounds with the same mass can be hard to tell apart without good chromatographic separation. Cost and throughput. The instrument is expensive, needs expertise, and is far slower than a plate reader, so it’s used for confirming top strains, not first-pass screening.

This is how the output looks:

The calculation can be obtained from the link below:

https://edit.htgaa.org/2026a-sheila-ramani/webpages/_edit/main/content/homework/week-10-hw-imaging-and-measurement/_index.md

3.For two of the mass spec instruments that you saw in the lab:

I chose the TQ Absolute and the BioAccord, because they fit two different measurements in my Yarrowia project: a small-molecule product (hydroxytyrosol) and the enzymes that make it (hpaB and hpaC).

What makes the instrument unique (or, what does it specialize in)?

It specializes in robust, high-sensitivity quantitation of analytes you already know you’re looking for, using two quadrupoles in tandem (the “TQ” in its name). It stays reliable across thousands of injections, and it’s used for clinical samples and for drug and metabolite quantitation in pharma.

What does the output data look like?

Not a spectrum, but an ion chromatogram: signal intensity against time for a selected precursor-to-fragment transition. Concentration comes from the peak area against a calibration curve.

Describe something you could measure for this class (drawing from previous labs or your project) using this instrument. What would the measurement help you understand?

Hydroxytyrosol and tyrosol in the culture broth and in cell extracts. I’d look for each compound by its mass and a fragment of it.This would show whether the pathway works, how much hydroxytyrosol the cells make, and how much leaks into the broth instead of staying inside.

What format would you need your sample in to perform this analysis (denatured, native, digested)? What other sample preparation might you need to do?

Sample format and preparation.

Small molecules, so denatured versus native doesn’t apply. Separate cells from broth by centrifugation. Filter and dilute the broth. For intracellular product, lyse the cells (bead beating or enzymatic) and extract with an organic solvent, then dry and reconstitute. Spike every sample with an internal standard, and run a calibration curve and QC samples. Hydroxytyrosol oxidizes easily, so keep samples cold and acidified, possibly with an antioxidant (my background knowledge; check it against a protocol).

4.Given these different measurement technologies, describe how you would use at least one instrument to measure something for your final project. Describe how you would prepare the sample, generate the data, what the data would tell you, and how it would advance your project.

Instrument: TQ Absolute (triple quadrupole LC-MS/MS)

What I’d measure: Hydroxytyrosol and its precursor tyrosol in my engineered Yarrowia lipolytica strains. Hydroxytyrosol is the least proven part of my design, since I found no report of it being made in this host. A plate-reader Nile Red screen can’t see it, so a targeted, sensitive method is needed.

Sample preparation

Cultures: Grow the engineered strain, the parent strain, and a strain with hpaB/hpaC but without the feedback-resistant ARO4, each in at least three replicates. Add a no-cell medium blank. Grow them under lipid-accumulation (nitrogen-limited) conditions, since that is when my design expects the pathway to be active. Broth: Centrifuge to remove cells, then filter and dilute the supernatant. Cells: Wash the pellet and lyse it (bead beating or enzymatic), extract with an organic solvent, then dry and reconstitute in the starting mobile phase. Standards and controls: Spike every sample with an internal standard, and run a calibration curve from pure hydroxytyrosol and tyrosol plus quality-control samples. Keep samples cold and acidified because hydroxytyrosol oxidizes easily (my background knowledge, so check it against a protocol).

Generating the data

Inject each sample into the LC, which separates the compounds on a reversed-phase column. The mass spectrometer selects each compound’s mass, breaks it up and monitors a fragment.The output is an ion chromatogram, and the peak area gives concentration against the calibration curve.

What the data would tell me

Whether the pathway works: a hydroxytyrosol peak at the right retention time and fragment, present in engineered strains and absent in the parent. How much is made: the concentration in the broth and in the cells. Where it ends up: the split between cells and broth, which affects the extraction step. Where the bottleneck is: tyrosol accumulating without hydroxytyrosol suggests the hpaBC step is limiting, while neither compound appearing suggests tyrosine or Ehrlich-pathway supply is limiting.

How it advances the project

It gives a go/no-go answer for the highest-risk module. If hydroxytyrosol is produced at a useful level, I’d carry it into the oil strain, and if not, I’d keep it as a separate stream made in another host and blended in. It also tells me which step to fix next, whether that’s enzyme expression or precursor supply, and it provides a benchmark against published titers in E. coli and S. cerevisiae. I’d set the numeric threshold after measuring the baseline, and use GC-MS and the plate reader for the oil and squalene.

AI Prompt:

What makes the instrument unique (or, what does it specialize in)?

What does the output data look like?

What format would you need your sample in to perform this analysis (denatured, native, digested)? What other sample preparation might you need to do?