Week 7 HW: Measurement & Assays
A. Analytical tools
Part 1 How is the sample introduced?
A liquid sample is injected into the LC system, where compounds are then separated on a column before entering the mass spectrometer. With electrospray ionization (ESI), the liquid forms charged droplets that lose solvent and release gas-phase ions. Samples can also be introduced by direct infusion.
Part 2 What physicochemical property is being leveraged?
The instrument separates ions with their mass to charge ratio (m/z) and uses the fragmentation behavior for additional selectivity. In multiple reaction monitoring (MRM), the first quadrupole selects a precursor ion, the collision cell fragments it, and the third quadrupole selects a specific product ion.
Part 3 What does this allow you to measure?
It allows sensitive, targeted measurement of compounds such as metabolites, drugs, and peptides. A selected precursor process ion transition supports identification, while its chromatographic peak area is compared with a calibration curve to determine concentration. Retention time and additional transitions increase confidence in identification; m/z alone does not uniquely identify a molecule.
Part 4 How is the output signal detected?
Selected ions reachs the instrument’s photomultiplier based XDR detector, which generates an amplified electrical response. The signal is digitized and processed by software to produce spectra or chromatograms showing ion intensity over time.
Part 5 What limitations does the tool have?
Compounds must ionize effectively to be detected. Other sample components can suppress or enhance ionization, affecting quantitative accuracy, so sample preparation and suitable standards are important. MRM methods require predefined target transitions and are therefore less suitable for discovering completely unknown compounds. The instrument also does not directly measure material properties such as stretchability or moisture response.
B. Liquid chromatography mass spectrometry (LC-MS) instrument
Part 1 How is the sample introduced into the LC? Into the MS?
A prepared liquid sample is placed in a vial, and an autosampler injects it into the LC’s flowing mobile phase. The sample passes through a column, which separates its components. The column outlet carries these components into the MS’s electrospray ionization (ESI) source, where charged droplets lose solvent and release gas-phase ions. BioAccord combines an ACQUITY UPLC system with an ACQUITY RDa time-of-flight mass detector.
Part 2 What physicochemical properties are being leveraged?
In a typical reversed-phase LC method, separation depends on differences in hydrophobicity: molecules interact differently with the stationary and mobile phases and therefore leave the column at different times. In the MS, ions are separated by mass-to-charge ratio (m/z). After acceleration, ions with lower m/z travel faster through the time-of-flight analyzer than those with higher m/z. Other LC methods can exploit different properties, such as molecular size.
Part 3 What does this allow you to measure?
BioAccord measures retention time, ion m/z, and signal intensity. Software can combine signals from different charge states to calculate an intact protein’s molecular mass. These measurements support protein mass confirmation, peptide mapping, and detection of mass changes associated with modifications or degradation. Quantification requires an appropriate calibration method; signal intensity alone is not concentration.
Part 4 How is the output signal detected?
Ions reach a detector after passing through the time-of-flight analyzer. Their arrival times and ion responses are converted into digitized signals. Software converts flight times into m/z values, producing a mass spectrum of intensity versus m/z. Tracking selected ions across the LC run produces an extracted ion chromatogram of intensity versus retention time.
Part 5 What limitations does the tool have?
Sample components and additives can suppress ionization, reducing sensitivity and affecting quantification. Fragile molecules can also fragment in the ion source. Intact mass alone cannot establish a complete protein sequence or uniquely locate a modification. Reversed-phase conditions can disrupt native protein structure, and the instrument does not directly measure material properties such as stretchability or moisture response.
C. Two mass spectrometry instruments
Cyclic IMS
Specialization: Cyclic IMS combines mass spectrometry with ion mobility, separating ions according to their movement through a gas under an electric field. Mobility depends on charge and molecular shape, allowing investigation of structural differences that mass measurements alone may not resolve.
Output: Mass spectra, chromatograms when coupled to LC, and mobilograms showing ion intensity versus mobility arrival time. Combined plots can show both m/z and mobility.
Class application: I could investigate whether soluble SELP forms different oligomers or conformational populations under different solution conditions. This could help explore its assembly behavior before fabrication. However, gas phase mobility would provide indirect structural evidence rather than a direct measurement of its structure in a hydrated material.
Sample preparation: I would use intact, undigested SELP under native compatible conditions, exchanging it into a volatile buffer such as ammonium acetate. I would remove particles and nonvolatile additives while checking that buffer exchange does not cause precipitation or substantially change assembly.
Xevo QTOF
Specialization: Xevo QTOF combines quadrupole ion selection with time of flight mass analysis. It supports accurate mass measurements and fragmentation analysis for investigating protein identity, sequence, and post-translational modifications.
Output: Mass spectra showing intensity versus m/z, chromatograms showing signal over retention time, and MS/MS spectra showing the fragments of selected precursor ions.
Class application: I could analyze peptides from my SELP product and compare them with peptides predicted from its designed sequence. This would provide evidence of protein identity and help investigate unexpected modifications or degradation.
Sample preparation: For peptide mapping, I would use denatured, enzymatically digested SELP, with reduction and alkylation if required by its sequence. I would select a digestion enzyme suitable for SELP’s repetitive sequence, remove incompatible salts and detergents, and clean up the peptides before LC-MS/MS.
D. Final Project
I think BioAccord would be helpful for my final proposal to check SELP mass and integrity before fabrication. After expression and purification, I would clarify the sample and remove incompatible salts or additives. I would keep SELP soluble and analyze it using a facility approved reversed phase LC-MS method, alongside a solvent blank and an instrument performance standard.
The LC would separate sample components, and the MS would generate mass spectra across the run. I would deconvolute the protein’s charge state signals and compare the measured mass with the theoretical mass of my exact construct, including any retained purification tag.
A dominant mass matching the expected SELP would support successful production of the intended full length protein. Additional lower mass species could suggest degradation or truncation, although further analysis would be needed to identify them. Intact mass alone would not prove the complete sequence.
This measurement would help me decide whether a protein batch is ready for fabrication or whether expression and purification need improvement. It would also help distinguish protein production problems from later fabrication problems. Stretchability and moisture response would require separate material tests.