Week 7 HW: Measurement & Assays

Part 1 — Fluorescence Plate Reader

  1. How is the sample introduced? Samples are added to a 96-well or 384-well plate using a pipette or a robot. Each well can contain a different cell-free protein synthesis (CFPS) reaction.

  2. What physicochemical property is being leveraged? Fluorescence. For example, sfGFP absorbs light at around 485 nm and emits light at around 510 nm.

  3. What does this allow you to measure? We can measure fluorescence intensity to estimate how much sfGFP is produced under different CFPS conditions.

  4. How is the output signal detected? The plate reader shines light into each well and measures how much fluorescent light is emitted. The results are shown as relative fluorescence units (RFU).

  5. What limitations does the tool have? Fluorescence intensity does not always reflect the actual amount of protein produced. Other factors, such as pH, protein folding, and background fluorescence, can affect the signal. This means two samples might show similar fluorescence levels even if they contain different amounts of protein.


Part 2 — Liquid Chromatography–Mass Spectrometry (LC-MS)

  1. How is the sample introduced into the LC and MS? The liquid sample is injected into the LC, where different molecules are separated as they pass through a column. The separated molecules then enter the MS, where they are ionized, usually using electrospray ionization (ESI).

  2. What physicochemical properties are being leveraged? LC separates molecules based on how they interact with the column, such as their hydrophobicity. MS separates ions based on their mass-to-charge ratio (m/z), which is the mass of an ion divided by its charge number.

  3. What does this allow you to measure? We can identify molecules based on their mass and measure their relative abundance. For proteins, LC-MS can help us identify peptides and check whether the expected protein was produced.

  4. How is the output signal detected? The MS detects ions at different m/z values and converts them into electrical signals. The results can be shown as a mass spectrum (m/z vs. intensity) or a chromatogram (signal intensity over time). A higher m/z means more mass per unit charge, not necessarily a heavier molecule.

  5. What limitations does the tool have? LC-MS usually requires sample preparation, and complex mixtures can make the results harder to interpret. Some molecules are also easier to ionize than others, so a stronger signal does not always mean there is more of that molecule. Accurate quantification often requires calibration.


Part 3 — Comparing Two Mass Spectrometry Instruments

For my project, the two instruments could answer different questions:

  • BioAccord: Did I produce the intended protein?
  • TQ Absolute: How much of that protein did I produce?

Instrument 1 — BioAccord

Accessibility text Accessibility text
  1. What makes this instrument unique? BioAccord uses LC-MS with a Time-of-Flight (TOF) analyzer. It is mainly used to identify proteins and measure their molecular mass.

  2. What does the output data look like? It produces a mass spectrum showing m/z against intensity. It can also show a chromatogram of the signal over time.

  3. What could you measure, and what would it tell you? For my final project, I could use BioAccord to check whether the elastin-like protein (ELP) produced in my cell-free system has the expected molecular mass. This would help me see whether I produced the full-length protein.

  4. What sample format and preparation would you need? I would use an intact protein sample, which could be denatured depending on the LC-MS method. I would first clean up the cell-free reaction mixture to remove other proteins, salts, and interfering substances. Since I want to measure the mass of the whole protein, I would not digest it into smaller peptides.

Instrument 2 — TQ Absolute

Accessibility text Accessibility text
  1. What makes this instrument unique? TQ Absolute is designed to measure very small amounts of known molecules. It is especially useful for accurately quantifying specific proteins or peptides.

  2. What does the output data look like? It produces ion chromatograms showing signal intensity over time. The area under a peak can be used for quantification.

  3. What could you measure, and what would it tell you? I could use TQ Absolute to quantify ELP production under different cell-free reaction conditions by measuring specific peptides from the target protein.

This would help me compare protein yields and identify which conditions are more efficient.

  1. What sample format and preparation would you need? I would denature and digest the protein into smaller peptides, then clean up the sample before LC-MS analysis. I would also need a suitable target peptide and calibration standards to measure the protein amount accurately.

Part 4 — Application to My Final Project

For my final project, I want to use E. coli lysate to produce elastin-like proteins (ELPs) as a potential alternative to spandex. I would use the TQ Absolute to measure how much ELP is produced under different cell-free reaction conditions.

First, I would collect samples from each reaction, digest the proteins into smaller peptides, and clean up the samples before LC-MS analysis. I would then use the TQ Absolute to measure a specific peptide from my target ELP and compare its signal with calibration standards to estimate the protein concentration.

Since ELPs have repetitive sequences, I could potentially use these repeated peptides to improve detection sensitivity. However, I would need to account for how many copies of the peptide are present in each protein and choose a suitable digestion method.

This would help me compare protein yields across different reaction conditions and identify which conditions work best.

Knowing how much protein I can produce would be an important first step before testing whether the material has the elasticity needed to replace spandex.