Week 7: Mass Spectrometry

Homework questions

Introduction

Goal: For this homework you will reflect on the different analytical tools you have used in class so far and integrate the learnings from the analytical module.

Analytical technologies are designed to measure different aspects of molecules and samples. All analyzers include the following components: 1) a mechanism to introduce the sample, 2) a mechanism to probe the sample, leveraging physiochemical properties of the target analytes, 3) a mechanism to detect the probed response (i.e. a detector/digitizer), and 4) a mechanism to analyze the data.

In this module, liquid chromatography and mass spectrometry were introduced as advanced tools for quantifying and characterizing the structure of large and small molecules. Mass spectrometers serve to separate and detect ions by their mass-to-charge (m/z) ratios, allowing identification by molecular weight. Virtually every analytical technique performs better with improved purity of sample; liquid chromatography can facilitate the specific detection of molecules by separating them in the time domain (prior to UV or MS detection) using properties such as the differential hydrophobicity, isoelectric point, or size of the molecules.

1. Analytical tools

For one of the analytical tools you have worked with thus far (e.g. plate readers, gel electrophoresis), describe:

a. How is the sample introduced?

In the class, we worked on plate readers to quantify the amount of sfGFP our CFPS systems produce. The samples are usually introduced to 96- or 384-well plates by pipetting, either by humans or robots.

b. What physiochemical property is being leveraged to study the sample with this tool?

For sfGFP, excitation light at a wavelength such as 485 nm excites the fluorophore, which then emits light at around 510 nm.

c. What does this allow you to measure?

This allows us to measure sfGFP fluorescence. By using a specific excitation wavelength and detecting light at a specific emission wavelength, we can reduce background interference.

d. How is the output signal detected?

The emitted light passes through an emission filter or monochromator and reaches a detector, usually a photomultiplier tube (PMT), which converts it into an electrical signal. By measuring the fluorescence of a series of purified GFP concentrations, we can build a standard curve and use it to calculate the GFP concentration corresponding to a fluorescence reading.

e. What limitations does the tool have?

The molecule needs suitable excitation and emission wavelengths. If its fluorescence overlaps with background from the sample or buffer, this method may have difficulty distinguishing it from the background.

The instrument and ambient light can affect the baseline, which may drift during measurement. Different instruments may also give different standard curves, so quantitative measurements depend on calibration.

2. Liquid chromatography mass spectrometry (LC-MS)

For a liquid chromatography mass spectrometry (LC-MS) instrument:

a. How is the sample introduced into the LC? Into the MS?

Into the LC: The sample is first dissolved in a suitable solvent, then introduced into the flowing mobile phase by an autosampler or a manual injector. A high-pressure pump pushes the mobile phase through an LC column containing the stationary phase, where the sample components are separated.

Into the MS: The eluate from the LC column enters the ion source. A common method, electrospray ionization (ESI), uses a high voltage to generate charged droplets. As the solvent evaporates, gas-phase ions form and enter the mass analyzer through the inlet and ion optics.

b. What physiochemical properties are being leveraged to study the sample with this tool?

LC separates molecules through differences in their interactions with the mobile phase and stationary phase, for example hydrophobicity, charge related to isoelectric point, or size of the molecules.

MS analyzes gas-phase ions by their mass-to-charge ratio (m/z). For example, Time-of-Flight (TOF) uses differences in ion flight times under defined conditions to determine m/z.

c. What does this allow you to measure?

LC-MS provides the retention time, m/z, and signal intensity of sample components, which help identify molecules. We can calculate the molecular mass by deconvoluting the charge states and m/z value. Integrating the signal for a target molecule allows quantitation. With MS/MS, it can also provide peptide sequence information and help identify and locate post-translational modifications (PTMs).

d. How is the output signal detected?

After passing through the mass analyzer, ions reach the detector. Common detectors, such as electron multipliers or microchannel plates (MCPs), amplify the electronic signal produced by ion impacts. Electronics then record the signal.

The software organizes these signals into a mass spectrum, with m/z on the x-axis and signal intensity on the y-axis, or a chromatogram, with retention time on the x-axis and signal intensity on the y-axis. For TOF, the system also records ion arrival times precisely and converts them into m/z through calibration.

e. What limitations does the tool have?

Molecules that are difficult to ionize may give weak signals or go undetected even when they are present.

Salts and detergents can also interfere with the analysis, so suitable sample purification is usually needed.

Different isomers can have the same molecular mass, so we need another way to distinguish them.

The instrument has a limit of detection (LOD), and a limited m/z range. Molecules at very low concentrations may be difficult to detect.

3. Mass spectrometry instruments

For two of the mass spec instruments that you saw in the lab:

a. TQ Absolute - Protein/peptide quantitation

TQ Absolute instrument. TQ Absolute instrument.

i. What makes the instrument unique (or, what does it specialize in)?

The TQ absolute is designed for robust quantitation of trace analytes. It requires the user first know the analytes they are looking for but features extremely high performance for thousands of injections. It is especially suitable for low-abundance proteins, peptides, or other products. It can zoom into the analytes that we are interested in. Good for analyzing a super low abundance protein or product. Waters Xevo TQ Absolute.

ii. What does the output data look like?

The TQ of the TQ Absolute stands for “Tandem Quad,” as the instrument features two quadrupoles that operate by selecting only ions of particular mass-to-charge ratios. In multiple reaction monitoring (MRM) mode, data is not recorded as a spectrum but as an ion trace over time, called an ion chromatogram. The x-axis is retention time, and the y-axis is the signal intensity of the selected precursor-to-product transition. Waters Xevo TQ Absolute guide.

Targeted ion chromatograms shown during the TQ Absolute demonstration. Targeted ion chromatograms shown during the TQ Absolute demonstration.

iii. Describe something you could measure for this class (drawing from previous labs or your project) using this instrument. What would the measurement help you understand?

It is often used for clinical samples (lab blood testing) and the quantification of drugs and drug metabolites during the DMPK phase of pharma discovery. I could also use it to detect low concentrations of histamine or peptides specific to sfGFP, but we first need to know the precursor and product ion m/z values. For some small-molecule drugs detection, I think this is a very good instrument. Waters, histamine quantification by UPLC-MS/MS.

iv. What format would you need your sample in to perform this analysis (denatured, native, digested)? What other sample preparation might you need to do?

For peptide detection, I would denature the protein and use trypsin digestion to convert it into peptides. Depending on the sample, protein cleanup, peptide enrichment, and desalting after digestion may also be needed.

For drug molecules or histamine, digestion is not needed. Depending on the sample, I may need to remove proteins, salts, and other interfering components before analysis.

b. Xevo QTOF - Post-translational modifications, proteoforms, quantitation

Xevo QTOF instrument and its mass spectrum display. Xevo QTOF instrument and its mass spectrum display.

i. What makes the instrument unique (or, what does it specialize in)?

Xevo QTOF is more suitable for broader characterization of a sample and can also be used for quantitation. It is ideal for probing the primary structure and PTMs of most therapeutics and performing quantitation. This would be something that gives a bigger picture of what’s in the sample. The tall tube at the back is the TOF analyzer.

ii. What does the output data look like?

The system features a TOF detector but also a device called a quadrupole (Q), which allows for a single m/z to be selected for deeper analysis. The QTOF generates mass spectra. QTOF stands for Quadrupole Time-of-Flight. The x-axis is m/z, and the y-axis is signal intensity.

For an intact protein, a group of peaks may come from the same protein at different charge states, as shown below. Deconvolution converts these peaks into the corresponding protein molecular mass (Da). Waters, routine intact protein LC-MS analysis.

Mass spectra showing a protein at different charge states. Mass spectra showing a protein at different charge states.

iii. Describe something you could measure for this class (drawing from previous labs or your project) using this instrument. What would the measurement help you understand?

I could analyze sfGFP produced by CFPS to check whether its intact mass matches the expected value and whether there are products with different masses, such as truncated products or some modifications.

iv. What format would you need your sample in to perform this analysis (denatured, native, digested)? What other sample preparation might you need to do?

My goal is to measure the intact mass, so I would enrich or purify sfGFP from the CFPS mixture, then desalt it to remove salts and detergents that are unsuitable for MS. I would keep the protein as native form.

4. Final project

Given these different measurement technologies, describe how you would use at least one instrument to measure something for your final project. Describe how you would prepare the sample, generate the data, what the data would tell you, and how it would advance your project.

For my final project, I want to make a test strip that detects histamine produced during food spoilage. I could use the TQ Absolute to measure the histamine concentration in a sample and compare it with fluorescence measured by a plate reader. This would help me check the relationship between the sensor fluorescence output and the actual histamine concentration.

Sample preparation

I would prepare a series of CFPS reactions with different histamine concentrations, including a control with no added histamine. After the reaction, I would split each sample into two portions: one for fluorescence measurement and the other for LC-MS. For the LC-MS/MS samples, I would use a suitable cleanup method to remove salts and other possible interferences.

Data generation

I would first use a histamine standard to confirm the retention time and select the precursor-to-product transition. The chromatogram would have retention time on the x-axis and the signal intensity of the selected transition on the y-axis. Waters, histamine quantification by UPLC-MS/MS.

After getting the data from histamine concentration measured by LC-MS and the sfGFP fluorescence measured by the plate reader, I would establish the relationship between histamine concentration and fluorescence, so I could use the fluorescence output to estimate histamine concentration in later sensor tests.

Appendix - other three mass spectrometry instruments

A) BioAccord - Protein sequence, protein mass

The BioAccord is designed for non-LC-MS users and features an LC for sample introduction and separation, self-diagnostic capabilities, and low-mid performance in terms of sensitivity. It is ideal for routine characterization problems (e.g. protein identification) and non-trace quantitative analyses and is commonly used for process analytical analyses in pharma. It uses a Time-of-Flight (TOF) analyzer that generates mass spectra (m/z vs. intensity plots). Individual ions can be plotted over time, which is referred to as an extracted ion chromatogram.

Related reference: Waters BioAccord system specifications.

BioAccord instrument during the lab tour. BioAccord instrument during the lab tour.

B) Cyclic IMS - Tertiary structure

The Cyclic IMS is a unique mass spectrometry system (QTOF) into which ion mobility is integrated into the flight path. This offers a unique tool set for probing the secondary and tertiary structure of proteins, lipids, and other molecules. It is ideal for deep characterization and across a range of applications; because of this, sample is often introduced by a variety of different methods, including LC, direct injection, or by 2D spatial imaging methods such as DESI (Desorption ElectroSpray Ionization). The addition of the ion mobility device adds a third dimension to the data and data can be viewed as a mobilogram (viewed in mobility space), a chromatogram (similar to BioAccord or Xevo, or as a mass spectrum. If an imaging technique is used, mass-to-charge spectra are recorded as a function of position, generating an image of ion intensities. This technique can be used to image drug and drug metabolite abundance in different tissues (e.g. liver).

Cyclic IMS instrument during the lab tour. Cyclic IMS instrument during the lab tour.Data displayed during the Cyclic IMS demonstration. Data displayed during the Cyclic IMS demonstration.

Related reference: Waters SELECT SERIES Cyclic IMS brochure.

C) CDMS - viruses, large multi-protein complexes, glycoproteins

The CDMS (Charge Detection Mass Spectrometer) is a special instrument used for the analysis of very large (megadalton) biomolecules in the folded (native state). It uses a unique detector that samples one ion at a time, allowing these very large ions to be properly measured. It is often used for AAV or LNP characterization in the pharmaceutical space. Data is often presented both as a traditional mass spectrum (m/z vs. intensity) and as a true mass vs. charge plot.

Very large molecules, like nanoparticles and AAV vectors, are in the megadalton range.

Gene therapy and cell therapy manufacturers need something to quality control their manufactured products as well.

This machine measures both m/z and charge, then uses them to calculate the exact mass.

Charge detection mass spectrometer during the lab tour. Charge detection mass spectrometer during the lab tour.CDMS data with a particle scatter plot and mass distributions. CDMS data with a particle scatter plot and mass distributions.

Plot: every single dot here is a particle.

Lower-left corner might be contaminants.

Use a syringe to push the sample in.

Related reference: Waters, large biomolecule analysis using CDMS.